even though you do not put test leads into input terminals
For better accuracy when measuring the dc offset of an ac
voltage, measure the ac voltage first. Note the ac voltage
range, then manually select a dc voltage range equal to or
higher than the ac range. This improves the accuracy of the
dc measurement by ensuring that the input protection circuits
are not activated.