MAINTENANCE
9.
Connect the 1 Ω standard and turn ON the R1L-D1.
Confirm the R1L-D1 automatically goes to the 2 Ω range,
and indicates 4-wire, an asterisk for auto-range, and the
resistance. Resistance values should also appear on the
PC screen.
10.
Type a lower case "r" and ENTER on the PC keyboard.
Verify a temperature sensor reading between 20 and 30
°C on the PC screen.
Step 2: Do not enter
a phone number.
Use COM port
selectable in the drop
down menu “Connect
using”
Step 3: Enter port
settings shown on
the right.