Operating instructions
– 43 –
6.
TROUBLESHOOTING
The steps to detect a faulty condition on the SERENITY NEXT chair are described in table-form. Each table reports the title of the
troubleshooting procedure, and three columns whose content can help to detect and solve the problem. Below is an example:
Procedure Name
Problem
Test Procedure
Corrective Action(s)
Patient chair problem description.
If the answer to the proposed question is
“YES” look at the question written in the
next box below; if the answer to the ques-
tion is “NO” take all the corrective ac-
tion(s) described in the right box before
moving to the next question.
Action(s) to be carried out in order to fulfil
the requirements of the question asked
by the test procedure at the side. Only af-
ter all actions have been carried out with
positive result it is possible to move to the
next test procedure.
The table-form takes into consideration only electrical or electronic faulty conditions as mechanical faults are easily detectable by
a simple visual inspection. In order to execute a visual inspection, it is sufficient to remove the upholstery parts, as described in
paragraph 5.4.
Main supply control procedure
No chair movement.
Is the supply voltage correctly rated?
If an OHM-meter or a similar testing de-
vice is not available, plug any other ap-
paratus (i.e. a lamp) into the socket and
verify the presence of the power supply.
Is the power cord plugged in?
Plug the cord into the socket.
Is the green light of the main switch on?
Turn the main switch on.
After having unplugged the chair, check
the main fuses. If necessary, replace
them as described in the user manual.
Otherwise call the technical service.
Does the PCB emit a double
“BEEP”
sound when the main switch is turned
on?
The PCB has probably failed. Call the
technical service.
The problem is not related to the chair
main supply group. Continue checking
the chair movements one by one.
Chair lifting movement control procedure
The chair does not lift.
Is there too much weight on the chair?
Apply the chair movement testing proce-
dure.
Remove the excessive weight. Look at
the chair operating instruction manual to
determine the maximum chair lifting ca-
pability.
Backrest tilting movement control procedure
The chair backrest does not tilt.
Is there an object that collides with the
chair, activating the safety device and
stopping the backrest tilt movement?
Apply the chair movement testing proce-
dure.
Remove the object that collides with the
chair.
Backrest raising movement control procedure
The chair backrest does not raise.
Was the chair movement testing proce-
dure duly applied?
Apply the chair movement testing proce-
dure.
Call the technical service.