16
7. STERILISATION TABLE
CYCLES
STERILISATION
TIME
DRYING TIME MAX. LOAD for
TRAY (Kg)
WORKING PRESSURE
(Bar Relative)
Not wrapped
121°C
20 min.
5 min.
1
1,2÷1,3
Not wrapped
134°C
10 min.
5 min.
1
2,1÷2,3
The data of this table are indicative; the choice of the sterilisation cycle must be based on the data
supplied by the producer of the object you want to sterilise.
8. FILLING THE WATER TANK
Remove the water tank cap, fill with distilled or demineralised water up to reaching the lower edge of the
sheet separation wall (see dwrg. D).
dwrg D
NOTE: The use of feedwater with contaminants at levels exceeding those given in the table above
can greatly shorten the working life of the device, with damage at the components of the autoclave
and then invalidate the manufacturer’s warranty.
TABLE OF QUALITY LEVEL IN ACCORDING TO DIRECTIVE DIN EN 285
CEN STANDARD DIN EN 285
Evaporation residue
≤
10
mg/l
Silicium oxide, SiO2
≤
1
mg/l
Iron
≤
0.2
mg/l
Cadmium
≤
0.005
mg/l
Lead
≤
0.05
mg/l
Rest of heavy metals except iron, cadmium, lead
≤
0.1
mg/l
Chloride (Cl’)
≤
2
mg/l
Phosphate (P20s)
≤
0.5
mg/l
Conductivity (at 20°C)
≤
15
µ
s/cm
PH value (degree of acidity)
5 to 7
Appearance
Colourless clean without sediment
Hardness (E Ions of alkalin earth)
≤
0.02
mmol/l