10 – MESSAGES OF ERROR OR ALARMS
Messages of error are emphasized through an Alfa-numerical code, consisting in a letter and three numbers.
The “E” suffix is relative to errors of the operator, which can be corrected by same operator.
The “A” suffix is relative to alarms, irregularities of the autoclave, after having carried out the remedy, if the
problem persists it is necessary to request telephonic assistance.
To reset alarms and errors, maintain pressed the buttons here indicated until the intervention of a
beep:
10.1 – LIST OF ERROR-ALARM MESSAGES
ERROR CAUSE
REMEDY
E 200
Low water level
Fill tank
E 250
Low level water cleaned up with loading
from deionizer
Fill tank
E 300
Maximum level used water
Empty tank
E 401
Door open
Close door with attention
E 451
Door unlocked
Open door and close again
E 461
High temperature in chamber for
Shut off and wait 10 minutes
Vacuum test (above 40°)
A 250
Cycle begins without water
Control Waterclean device
A 001
Interrupted cycle
Reset and restart cycle
A 011
Irregular pc display
Shut off and restart autoclave
A 101
Vacuum not reached in 10min.
Reset and repeat the cycle
A 111
Vacuum not maintained for first phase VT
Reset and repeat the cycle
A 121
Vacuum not maintained for second phase VT
Reset and repeat the cycle
A 401
Hatch opened during the cycle or problem closing To control the closing system
A 501
Pressure not maintained
Reset and repeat the cycle
A 551
Irregular 1° temperature sensor
Reset and repeat the cycle
A 552
Irregular 2° temperature sensor
Reset and repeat the cycle
A 641
Overly high pressure
Reset and repeat the cycle
A 651
Overly high temperature
Reset and repeat the cycle
A 661
Irregular temperature comparison
Reset and repeat the cycle
A 701
Working pressare not reached
Reset and repeat the cycle
A 751
Low temperature
Reset and repeat the cycle
A 761
Irregular pressure sensor
Reset and repeat the cycle
A 801
Irregular unloading pressure
Reset and repeat the cycle
A 901
Main voltage too low
Inform electric energy provider
SUBSTUTUTE RESIN CLEN WATER resin finished
Follow the instruction in Paragraph 14.2 of the
manual
D
The printer’s door is not closed
Try to close the printer’s door
P
Paper is missing
Insert paper
L
There is low voltege on the line
To try to connect the machine to the net
electrical worker in an other point; if the
problem persists to contact a electrician
Summary of Contents for Europa B xp
Page 31: ...31 DXBA408 Rev_14 SCHEMA DI COLLEGAMENTO DEL DEMINERALIZZATORE VALIDO FINO AL 11 05...
Page 32: ...32 DXBA408 Rev_14 SCHEMA DI COLLEGAMENTO DEL DEMINERALIZZATORE VALIDO DAL 12 05...
Page 67: ...67 DXBA408 Rev_14 CHEME OF CONNECTION OF THE VALID DEMINERALIZER UNTIL THE 11 05...
Page 68: ...68 DXBA408 Rev_14 SCHEME OF CONNECTION OF THE VALID DEMINERALIZER FROM THE 12 05...
Page 104: ...104 DXBA408 Rev_14 SCHEMA DE BRANCHEMENT DU DEIONISATEUR VALIDE JUSQU AU 11 05...
Page 105: ...105 DXBA408 Rev_14 SCHEMA DE BRANCHEMENT DU DEIONISATEUR VALIDE DU 12 05...
Page 141: ...141 DXBA408 Rev_14 ESQUEMA DE ENLACES DEL DESMINERALIZADOR V LIDO HASTA EL 11 05...
Page 142: ...142 DXBA408 Rev_14 ESQUEMA DE ENLACES DEL DESMINERALIZADOR V LIDO DEL 12 05...
Page 178: ...178 DXBA408 Rev_14 ANSCHLUSSCHEMA DES G LTIGEN ENTZSALZERS BIS DAS 11 05...
Page 179: ...179 DXBA408 Rev_14 ANSCHLUSSCHEMA DES G LTIGEN ENTZSALZERS VOM 12 05...