background image

IST-2137.EO01.01/A

 

Istruzione / User’s Manual / Manuel d’utilisation 

Pag.6/11

 

 

 

 

TECNOCONTROL S.r.l. Via Miglioli 47 SEGRATE ( MI ) 

 Tel: 02/26 92 28 90   Fax: 02/21 33 734

 

 

Sensor operating limits: 

the electro-chemical cell, in any case, should not be exposed to temperatures lower 

than -40° or higher than 60°C. The electrochemical cell should not be exposed to an over-range concentration of 
30 %  O

2

. If this condition occurs, we suggest replacing the sensor. 

Safety:

 Oxygen sensors contain a small quantity potassium acetate solution which is corrosive. Normally this ma-

terial would only leak out as a result of either mechanical damage or an incorrect instrument use beyond the limits 
of its operational functioning. These sensors also contain small amounts of lead, lead oxide, platinum, silver, car-
bon

 

and antimony. If the user comes into contact with the oxygen sensor contents the affected area should be 

washed with a copious supply of water. 

Cross sensitivity to other gases:

 The electrochemical cell employed has a good resistance towards products 

such as sprays, detergents, ammonia, glues and paints. 

However, in case of products containing substances in 

great quantity, these could interfere with the Sensor and cause false alarms.

 We recommend ventilating the room 

when products like these are used. Volatile organic solvents in great quantity cause false alarms, and then the cell 
needs one or more hours in clean air to recover the normal sensitivity conditions. 

TEST and CALIBRATION 

PAY ATTENTION

: This procedure has to be made with extreme attention and by authorized and trained people; 

because starting this procedure it will start both Outputs (relays) causing the activation of connected alarm devices. 
The instrument has two different code protected functions: 

Operation Check and Calibration

.  

Operation Check, Calibration in Air and Fine mA ADJ Output :

 are different code protected functions. To ac-

cess these functions is necessary to insert the relevant “

Code

" through the keys F1s and F2. To have the key 

pressure recognized, hold pressing it for around a second (

until the Green Led doesn't switch off for a moment

). Then the 

next key can be pressed. In case of error all it takes is waiting around 10 seconds and the sequence is automati-
cally erased. The 

Calibration Check

 is a function code free. 

Calibration Kit, Sample Gas Bottles

 (

for Calibration Check

please, using a mixture about 16%v/v O

2

 in Nitro-

gen.

 It is possible to use either the disposable cylinders with adjust valve or the high pressure ones with reduction 

gear. Is also necessary the Tecnocontrol Calibration Kit model TC011. 

“INSTRUMENT OPERATION CHECK” (

Check Code

: F2, F2, F1, F1)

this function allows to effect a functional test 

of the equipment. After having put the system in safety and inserted the "

Code Test

", all Led are switched off and all 

the relays are disarmed. Then they will switch on in sequence, the Led, from the yellow up to the 3° red. To the 
lighting of the various Led it corresponds the activation of the relevant relays (

relay "FAULT" with the yellow Led, relay 

"ALARM1" with the 1° red Led, relay "ALARM2" with the 2° red Led, relay "ALARM3" with the 3° red Led

). At the end all the Led will 

remain lighted for around 5 seconds, then the central returns at the conditions of normal operation. It is advisable 
to perform this operation every 6-12 months according to the use. 

Note: this function is not working if the 1° and/or the 2° and/or the 3° red Led have already turned on  

“CALIBRATION IN AIR” (

Calibration Code

: F2, F2, F2, F1, F2, F1)

this function allows completely recalibrate the 

sensor at 20,9%v/v in air. This procedure when it is launched, will automatically until the end. 

Warning

: to guarantee that no errors of elaboration happen, the rare possibility exists that during the Calibration 

the 

yellow LED switch off every 8 seconds, 

in this case interrupt the procedure, switch off and witch on the instru-

ment and repeat the Calibration. If condition persists it will be necessary to send the detector to the supplier for the 
reparation. 

Important note:

 

During Calibration routine the mA output indicates 0mA

.

 

The Calibration can be done in clean air only (

environment without the presence of flammable or other polluting gas

). With the 

keys perform the “

Calibration Code

 ". Wait until the Yellow and Green Led switch on fix and the 1

st

 red Led starts to 

flash (the 3

th

 Led Red

 

switch on to indicate the calibration working). Wait while the 1

st

 red Led switched off. At this 

point we can have two possibilities: 

Yellow and Green Led illuminates:

 the calibration routine has correctly been performed. The detector automatically 

switch off in 8 seconds, and restart into normal operating mode. (

see chapter OPERATIONAL DESCRIPTION “Preheating”

). 

Yellow LED illuminates:

 the calibration routine has failed. The detector automatically switch off and restart in 8 

seconds and repeat the procedure without inserting again the sequence. If condition still persists after the re-
placement of the cartridge it will be necessary to send the detector back to the manufacturer for reparation. 

Important

: If the environment is very polluted and there is doubt that the concentration of oxygen is 20.9%v/v, you 

can use an air cylinder with 

20.9%v/v of O

2

 or use a gas balloon (our code PA020 ) filled with fresh air. In this 

case, before executing the "

Calibrating Code

" insert the TC011 over the sensor holder, adjust the flow of air, so that 

the flow meter indicates approximately 0.3 l / min (

see Fig.3

). 

“CALIBRATION CHECK” (

no Code required

):

 

although the sensor operation is guaranteed if it is regularly per-

formed successfully "Calibrating in Air", it is possible to simulate the response of the detector using sample gas cyl-
inder with a mixture 16%v/v Oxygen in nitrogen. Insert the TC011 over the sensor holder; adjust the sample gas 
cylinder valve as the flow meter indicates around 0.3 l/mins (

see Fig.3

), wait for 1÷2 minutes or until the instrument 

activates Alarms as shown in chapter “

Operational Description

”. 

If the result is different, is necessary to recalibrate 

the sensor (

see “Calibration in Air”

).

 Then, close the gas cylinder

 

and remove TC011, 

wait 2÷3 minutes, until the in-

strument restores the normal working conditions

NOTE: further to the above, if the mA output is also used, verify with a voltmeter connected to the Test-Point 

TESTmA

” (

see Fig.2

), the value reach a 174mV value in normal clean air condition (20.9%v/v). 

[or if the mA output is not used, with the milliamperometer probe, directly on the terminals "S and -" verify that the 
value is about 17.4 mA]. Applying the 16%v/v O

2

 gas verify to read a value of about 143 mV. [or that the mA output 

decrease to about 14.3 mA]. Then the values, after removing the gas gradually returning to normal value in air. 

Summary of Contents for SE137EO

Page 1: ...e Time Temps de réponse T90 20 secondi seconds secondes Ripetibilità Repeatability Répétitivité 1 del segnale signal Precisione Accuracy Precision 2 del segnale in aria signal in air du signal en air Deriva a lungo termine Long time drift Dérive à long terme 2 segnale mese signal per month du signal chaque mois 5 sulla vita operativa over Operating Life sur sa vie Temp umidità di funzionamento Ope...

Page 2: ...ndi il Led giallo si spegne e si accende il Led verde ON che indica il normale funzionamento Dopo questo tempo il sensore è in grado di rilevare il gas ma raggiunge le condizioni di stabilità ottimali dopo circa 15 minuti di funzionamento continuo Funzionamento Normale deve essere acceso il solo Led Verde ON ALLARMI i tre livelli d allarme possono essere impostati con diverse soglie utilizzando i ...

Page 3: ...trochimici per Ossigeno contengono una piccola quantità di acetato di potassio che è corrosivo Normalmente non si verificano perdite dal sensore questo può essere causato solo da un azione mec canica violenta o un uso oltre i limiti di funzionamento Il sensore contiene anche piccolo quantità di piombo ossido di piombo platino argento carbonio e antimonio Nel caso avvenga un contatto del liquido de...

Page 4: ...i circa 136mV ovvero che l uscita in mA diminuisca fino a circa 13 6 mA Poi i valori dopo aver tolto il gas torneranno progressivamente al valore normale in aria Regolazione Fine Uscita in mA Codice Reg Fine F2 F1 F1 F2 questa funzione permette di regolare l uscita in mA di 0 32mA corrispondente a 0 5 O2 ma va utilizzata solo in fase di installazione e se dopo aver fatto la CALIBRAZIONE IN ARIA pu...

Page 5: ...t try to perform the procedure of Calibration in Air as described in the section Test and Calibration Calibration in Air finally try to replace a new Cartridge Sensor If the condition is not change it will be necessary to replace the unit and or send it back to the supplier to repair All LED activate FAULT relays activate and 24mA output signal this happens when the Cartridge Sensor is not working...

Page 6: ...l operation It is advisable to perform this operation every 6 12 months according to the use Note this function is not working if the 1 and or the 2 and or the 3 red Led have already turned on CALIBRATION IN AIR Calibration Code F2 F2 F2 F1 F2 F1 this function allows completely recalibrate the sensor at 20 9 v v in air This procedure when it is launched will automatically until the end Warning to ...

Page 7: ... mais étant fortement oxydant il réagit avec les matériaux combustibles et réduc teurs en facilitant leur combustion déterminant ainsi des risques d incendie et d explosion L Oxygène constitue 20 95 de l air que nous respirons Si la concentration d O2 dans l air descend sous 19 5 l on considère que cette carence est dommageable pour la santé des personnes si cette concentration descend sous 16 ce ...

Page 8: ...entation de plusieurs détecteurs en parallèle il est nécessaire de calculer la chute de tension sur les portions communes des câbles Un câble sans écran est normalement utilisé avec 2 conducteurs pour l alimentation les conducteurs pour les sorties relais Dans le cas ou l on utilise également le signal de sortie en 4 20 mA il convient d utiliser du câbles à écran à 3 conducteurs les conducteurs po...

Page 9: ...r rompre la procédure puis exécutez le contrôle en coupant et en rétablissant l alimentation du détecteur En cas de persistance il sera nécessaire de remplacer et ou renvoyer le détecteur au fournisseur pour réparation Important Durant l opération de étalonnage l appareil portant la sortie S au 0mA L étalonnage doit être exécuté en air propre exclusivement milieu sans la présence de gaz toxiques o...

Page 10: ...ogni rivelatore dall alimentatore a 12Vcc The maximum distance to install each detec tor from the 12Vdc power Supply Distance max d installation d un détecteur sous 12Vcc La max distanza cui può essere installato ogni rivelatore dall alimentatore a 24Vcc The maximum distance to install each detector from the 24Vdc power Supply Distance max d installation d un détecteur sous 24Vcc 0 75 mm2 26 Ω km ...

Page 11: ...ivelazione gas Gas Central Unit Centrale de détection RL Uscita in 4 20 mA 4 20 mA Output Sortie en 4 20 mA Dip Switch Test Uscita Output Test Sortie Test 40 200mV Fig 3 Tester di calibrazione Calibration Tester Kit de Calibration 0 3L min F1 F2 Tasti Key Touches Alimentazione Power Supply Alimentation 100 300 400 500 50 0 5 200 Vdc RL Resistenza di carico Load resistor Resistance de charge RL Ohm...

Reviews: