Heat meter
34
Parameters
Frost protection
(shown only if the Spe
-
cific heat capacity input
variable is
unused)
Specification of the antifreeze component in %
Reversing lock
Available for selection:
Yes
/
No
Status
Calibration value
Display:
Not calibrated
or
Calibrated
Display of the differential T.flow – T.return measured during the cali
-
bration process (in
Not calibrated
status this value must be 0.0 K)
Start calibration
Select to start the calibration process (Note the section
Calibration
process
!)
Delete calib. values
This can be used to
undo
the calibration. It resets the calibration
value to 0.
Delete meter reading
Button for deleting all meter readings
•
Antifreeze component (frost protection):
An average has been calculated from the product
specifications of all the major manufacturers and tabulated in relation to the mixing ratio.
Under typical conditions this method results in an additional error of
no more than
one per
-
cent.
•
Reversing lock:
Entering
No
permits
negative
metering; entering
Yes
means that the heat
meter will only be able to meter
positive
values.
• The tolerance of sensors and measuring device can sometimes lead to substantial errors in
the calculation of the differential temperature. The device has a
calibration process
to com
-
pensate this error.
• After selecting
Start calibration,
the user is prompted to confirm. If the calibration was per
-
formed by mistake or incorrectly, the result can be undone with
Delete calib. values
and/or
corrected by means of a new calibration.
Calibration process
By simultaneously measuring the
same
temperature with both sensors, the deviation of the sen
-
sors from each other can be ascertained and included as a correction factor in future calcula
-
tions.
The calibration affects only the sensor values in the Heat meter function and is not factored into
other functions.
During the calibration process, it is very important that both sensors (flow and return) capture the
same temperatures. This is achieved by binding the two sensor tips together with a piece of tape
or wire. Both sensors should also be fitted now with the lead extensions which will be used later,
so that the electrical resistances of the leads are included. If the collector sensor is being used,
the required lead length should be estimated and integrated. The sensors must be connected to
the two
programmed
inputs for flow and return and are immersed together in hot water (so that
both are exposed to the same temperature).
Calibration process:
1. Immerse the sensors in the water.
2. Start the calibration process and confirm.
Status display:
Calibrated
.
3. The calibration values displayed in the parameters and the corrected return temperature is
included in the output variables.
Summary of Contents for CAN-EZ2
Page 5: ......
Page 62: ...62 Messages This menu item displays activated messages Example A message is active...
Page 98: ......
Page 99: ......