Problem
Probable cause
Solution
The displayed flow rate
is higher than expected
The electrodes are immersed but
the pipe is not completely full
Make sure that the pipe is
completely full, for example,
installing the flowmeter in a vertical
pipe with upwards flow.
Blown fuse
Change the fuse
Not enough current from the power
supply
Change the power supply
The analog output gives
always 4 mA or 20 mA
Current output range not properly
programmed
Program the range properly (see
page 39)
The analog output gives
always 0 mA
Cable disconnected
Check the cable connection
The invalid current
message appears
Insufficient current from the power
supply
Change the power supply
Display is blank