13
As shown in the following table, to know the distance, the thickness of the pipe
s
must be
known.
In order to make the positioning of the pipe adaptor in the pipe easier, there is a label on one
of its sides with markings indicating the position of the internal pipe diameter for each DN.
Cut this label above the line corresponding to the DN of the pipe, at a distance equal to the
pipe thickness. Peel off the bottom part of the label. Once the pipe adaptor is placed into its
final position, where the label was cut must coincide with the outer diameter of the pipe.
This ensures that the measuring electrodes penetrate far enough in the area of flow profile
that will allow an accurate measurement.
DN
C (mm)
Insert pipe adaptor
L (mm)
H (mm)
80
10
93
88-s
100
12.5
85.5-s
125
15.5
82.5-s
150
19
79-s
200
25
73-s
250
31
67-s
300
37.5
60.5-s
350
44
54-s
400
50
48-s
500
62,5
145
140.5-s
600
75
128-s
700
87.5
115.5-s
800
100
103-s
900
112.5
90.5-s
1000
125
78-s
1200
150
203-s
1400
175
178-s
1600
200
153-s
1800
225
128-s
2000
250
103-s
205
450
56
42-s