8. Cleaning & Maintenance
8.5.2 Disinfection
Procedure
The following procedure should be used to disinfect the outside surfaces of the
instrument.
1. Disconnect the instrument from the mains power supply.
2. Disconnect the instrument from any accessories that are used.
Ensure that you are wearing disposable gloves.
3. Carefully wipe all the outside surfaces of the instrument with a wad of cotton
wool that has been soaked in the disinfecting solution.
4. Ensure that the same disinfection procedure is performed with the plate
carrier.
5. Repeat the disinfection procedure on any accessories which are also being
moved or returned.
6. After the disinfection procedure has been performed, ensure that the
disinfection certificate is completed.
See 8.5 for an example of the disinfection certificate that must be completed
before the instrument is returned to the distributor for servicing.
June 2002
Operating Manual for GENios, GENios FL, and GENios
Plus
No: I 112 904 Rev No: 1.1
8-5