8. Cleaning & Maintenance
8.4 Maintenance
CAUTION
Ensure that the instrument positioning is checked
using the service tool plate once a year.
8.4.1 Daily
No daily maintenance is necessary.
8.4.2 Weekly
Clean the cover and the transport with a mild detergent.
8.4.3 Every
Six
Months
Clean the filters using an optical cleaning solution (Lens tissue recommended).
8.4.4 Yearly
Maintenance
Check the belts tension (between 50 an 60 Hz, tension meter necessary).
Perform a FlouCheck test.
June 2002
Operating Manual for GENios, GENios FL, and GENios
Plus
No: I 112 904 Rev No: 1.1
8-3