5. Instrument Features
5. Instrument Features
5.1 Specifications
(GENios)
The following types of measurement are provided with the GENios microplate
reader.
Measurement Type
Read Mode
Microplates to be Used
Fluorescence (see 5.1.1) Standard and time-
resolved, top and bottom
6 - 384 wells
Absorbance (see 5.1.2)
Single and dual
wavelength
6 - 384 wells
Glow type luminescence
(see 5.1.3)
96
wells
The instrument allows for kinetic measurements
.
Reading may be restricted to one part of the microplate.
Up to four-measurement parameter sets can be queued using
XF
LUOR
.
Use of cuvettes
Please Note:
For cuvette use TECAN provides an adapter plate which fits up to 4
standard 10 mm cuvettes that can be measured simultaneously
(compatible with Hellma 10 mm 110-QS cuvettes equipped with PTFE
plug).
The cuvettes are positioned horizontally in the 4 designated areas.
The cuvette should be filled completely and closed tightly in order to exclude air
from the measurement chamber and ensure reliablity of results.
Select the provided CUV.pdf plate format from the plate definition file before the
measurement is started.
The table below lists the technical specifications for the instrument.
Parameters Characteristics
Measurement Software
controlled
Interface
Serial interface: RS 232
Filter handling
External filter exchange
Up to 4 excitation filters and 4 emission filters in
separate filter slides.
Up to 8 excitation and emission filter slides can be
defined when using software.
Plate definition
Via scanning software
Temperature control
From a5°C up to 42°C
Plate shaking
Orbital or linear shaking - both offer 3 speeds
June 2002
Operating Manual for GENios, GENios FL, and GENios
Plus
No: I 112 904 Rev No: 1.1
5-1