8. Cleaning & Maintenance
8.3 Replacing the Fuse
•
Turn the instrument OFF and unplug the mains cable from the instrument.
The fuse is located on the right hand side of the mains cable connection in
the rear of the instrument.
•
Open the fuse compartment by inserting a screwdriver into the slot in the top
of the cover and push the fuse holder out.
On/Off Switch
0 = ON
1 = OFF
Fuse compartment
Mains power socket
•
Pull the fuse holder out and replace the defective fuse(s) with the spare
fuse(s).
•
Ensure that the fuse(s) has/have the correct rating.
•
T 2 / 250V (slow blow)
Caution
Risk of Fire
For continued protection replace only with the
same type and rating of fuse.
•
Replace the fuse holder.
•
Reconnect the mains cable.
•
Turn the instrument on.
If the fuse continues to blow, call for service.
8-2
Operating Manual for GENios, GENios FL, and GENios
Plus
No: I 112 904 Rev No: 1.1
June 2002