3 - Software Communication
Communication Protocols
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
3 - 15
Example 1. The pump is set to address 0
Pump sends:
Dir
Group
Device
Frame
RTR
Length
1
001
0000
010
0
0000
Host acknowledges:
Dir
Group
Device
Frame
RTR
Length
Node ID
Slave ID
0
001
0000
000
0
0010
0010 0000
0010 0000
Host acknowledges the boot request with:
Dir = 0
Host to slave
Note
:
Boot MID is the same for
all nodes
Group = 1
Boot request response group
Device = 0
Always 0 in boot response
Frame = 0
Boot request response frame
RTR = 0
Always 0
Length = 2
Two data bytes in return message
Node ID
Group ID (2) +
Pump Address (0)
“ ”
00h
Must respond with
Group & Address
Slave ID
Same as Node ID
(hex 20)
“ ”
00h