E - 2
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
E - XLP 6000 Physical Specifications
Valves
Plug Material
Virgin Teflon
Body Material
Kel-F
Fittings
1/4-28 and M6 tubing fittings
1/4-28" syringe fitting
Valve Options
120° 3-port
90° 4-port
3-port distribution
Y block
T valve
6-port distribution
9-port distribution
Fluid Contact
Glass, Kel-F, Teflon, ceramic
Power Requirements
Voltage
24V DC + 10%
Current
1 A (peak)
Interface
Type
RS-232, RS-485 and CAN
Baud Rate
9600 or 38400 (RS-232 and RS-485)
100K, 125K, 250K, 500K and 1M (CAN)
Format
Data bits: 8
Parity: No
Stop bit: 1
Half duplex
Communications
Addressing
Up to 15 individual addresses available
Communications
Data terminal and OEM protocol (with error recognition)
Firmware
Programmable ramps
Programmable plunger speeds
Cutoff speed
Backlash compensation
Programmable delays and loops
Change speed on the fly
Terminate moves
Diagnostics
Absolute or relative positions
Programmable non-volatile memory