F - CAN Communication Commands
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
F - 1
F CAN Communication Commands
Command Type
Command
Valid/Invalid
CAN Equivalent
Initialization
Z, Y, W
Valid
None
Initialization
z
Valid
None
Plunger Movement
A, a, P, p, D, d
Valid
None
Valve
I, O, B
Valid
None
Valve
E, ^
Valid
None
Set
S, V, v, c, L, K, k
Valid
None
Command for
microstep-enabled
firmware
N
Valid
None
Control
G, g, M, H
Valid
None
Control
X
Valid
Frame type = 2
Command = “3” (ASCII)
Control
R
Valid
Frame type = 2
Command = “1” (ASCII)
Control
T
Valid
Frame type = 2
Command = “4” (ASCII)
Control
Reset
Invalid
Frame type = 2
Command = “0” (ASCII)
Control
Clear loaded command
Invalid
Frame type = 2
Command = “2” (ASCII)
Control
J, s, e, U
Valid
None
Report
?0 through ?29
Invalid
Frame type = 6
Command = “0” through
“29” (ASCII)
Report
F
Invalid
Frame type = 6
Command = “10” (ASCII)
Report
&
Invalid
Frame type = 6
Command = “23” (ASCII)
Report
Q
Invalid
Frame type = 6
Command = “29” (ASCII)