B - Plunger Information
Plunger Time Calculations
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
B - 5
Total Move Time
Ramp Up increments
Case 4. Move Too Small to Reach Top Speed
Case 4 is used when Vn < V and Vn > c
Diagram of Move:
Calculation:
Top speeds
Total Move Time
seconds
full increments
v = 50 Hz
L = 14
V = 5800 Hz
A = 350 full increments
c = 900 Hz
Hz
Hz
t
Vn
v
L
=
−
=
−
∗
=
838 50
14 2500
023
.
A
A
1
5
=
=
v
V
Speed [Hz]
c
time in s
t
Vn
Vn
AL
v
c
=
+
+
=
∗
∗ ∗
+
+
=
2
2
2 350 14 2500
50
900
2
4991
2
2
2
2
t
L
Vn
v
c
L
=
− −
=
∗
−
−
=
1
2
1
2 4991 50 900
26
(
)
(
) .