Alpha 1000 Handbook
Special Instructions for medical applications (IEC 60601-1:1988 + A1:1991 + A2:1995; EN 60601-1: 1990 +
A1:1993 + A2:1995; UL 60601-1, 1st Edition, 2006-04-26 (includes National Differences for USA) CAN/CSA-
C22.2 No. 601.1-M90 (R2005) (includes National Differences for Canada); IEC 60601-1:2005; EN 60601-1:2006;
ANSI/AAMI ES 60601-1:2005; CSA 22.2 No 60601-1:2008) - Applicable to products with L, R and T Input
Filter Options only
These products are designed for continuous operation within an overall enclosure, and must be mounted such that
access to the mains terminals is restricted.
These products are NOT suitable for use in the presence of flammable anaesthetic mixtures with air or with oxygen,
or with nitrous oxide.
For IEC60601-1, EN60601-1 and UL/CSA 60601-1 (Second Edition standards), the Alpha 1000 range provides
Basic insulation between input and outputs, and functional insulation between outputs and earth.
For IEC 60601-1:2005; EN 60601-1:2006; ANSI/AAMI ES 60601-1:2005; CSA 22.2 No 60601-1:2008), the Alpha
1000 range provides reinforced insulation between input and outputs (2 MOOPs) for all products. Between outputs
and earth there is functional insulation only.
Connect only apparatus complying with the standards listed above as appropriate to the signal ports.
These products are classed as ordinary equipment and are not protected against the ingress of water (IPX0).
Reference should be made to local regulations concerning the disposal of these products at the end of their useful
life.
Where any part of this product is made accessible to the operator in the end use equipment, the operator must not
touch this part and the patient at the same time.
These products have not been assessed to IEC/EN60601-1-2 (EMC) but EMC test data is available from TDK-
Lambda UK Ltd.
WARNING: No modification of this product is allowed.
Except for permanently installed equipment, the overall equipment into which these products are installed must have
double pole fusing on the input mains supply. The products themselves have single pole fusing in the live line only.
Environmental Specifications:
Description
Operation
Storage
Use
Indoor
-
Temperature
0 to 50°C (derating 2.5%°C above 50°C to 65°C -
Not covered by approvals).
-40°C to +70°C
Humidity
5 to 95% RH non-condensing
5% to 95% RH non-
condensing
Altitude
-200m to 3000m
-200m to 5000m.
Pressure
70kPa to 106kPa
54kPa to 106kPa
Orientation
Horizontal with chassis lowest, on either side or
vertical with the airflow upwards.
All
Material Group
IIIb
Pollution Degree
2
Overvoltage Category
II
Class
I
Weight
3 kg (Dependent on configuration)
IP Rating
IPX0
Level of Insulation:
Dielectric Strength testing is carried out as follows:
Primary mains circuit to earth - 2.25 - 2.35kVDC
Primary mains circuits to transformer core - 4.25 - 4.35kVDC*
Primary mains circuits to secondary -4.25 - 4.35kVDC.*
Outputs to each other and to earth are isolated to 500VDC.
*This test is not possible with modules fitted to the unit as damage to RFI capacitors will occur)
Summary of Contents for Alpha 1000
Page 20: ...Page 20 of 29 17027 issue 23 June 2014 Alpha 1000 Handbook Main Base PCB...
Page 22: ...Page 22 of 29 17027 issue 23 June 2014 Alpha 1000 Handbook Single Module PCB Dual Module PCB...
Page 23: ...Page 23 of 29 17027 issue 23 June 2014 Alpha 1000 Handbook Twin Module PCB J Module PCB...
Page 24: ...Page 24 of 29 17027 issue 23 June 2014 Alpha 1000 Handbook L Module PCB S Module PCB...
Page 25: ...Page 25 of 29 17027 issue 23 June 2014 Alpha 1000 Handbook T Module PCB MF Option MFV Option...
Page 26: ...Page 26 of 29 17027 issue 23 June 2014 Alpha 1000 Handbook...