• Use a test strip immediately after removing it from the tube.
• On the label, write the date you first opened the tube. Dispose of any remaining test
strips after 3 months.
• Do not use a test strip after its expiry date, as the result may be inaccurate.
• Do not bend, cut or alter the blood sugar test strip.
• Keep tubes and foil packaging containing test strips out of reach of children, as the
cap and test strips could be a choking hazard. Immediately alert the emergency
services if any part is swallowed.
For further information, see the test strips package insert.
Important information on the control solution
• Only use our control solution for your meter.
• Do not use the control solution after its expiry date or more than 3 months after
opening it. Write the date you opened it on the control solution bottle and dispose
of any remaining solution after 3 months.
• The recommended room temperature when performing a test with the control
solution is between 20 °C and 25 °C (68 °F and 77 °F). Ensure that the control
solution, your meter and the test strips are at the recommended temperature
before conducting a test.
• Shake the control solution well before use. Discard the first droplet of control
solution and wipe the dispensing tip clean in order to guarantee a pure sample
and accurate result.
• Keep the control solution properly sealed at a temperature between 2 °C and 30 °C
(35.6 °F and 86 °F).
DO NOT
freeze.
If you perform the recommended action but the issue continues, please call your local
customer support services.
TROUBLESHOOTING SYSTEM ISSUES
Reading the results
MESSAGE
WHAT IT MEANS
or 1,6 mmol/L
1.6 mmol/L (29 mg/dL)
KETONE?
13.3 mmol/L (240 mg/dL)
> 33.3 mmol/L (600 mg/dL)
Foutmeldingen
MESSAGE
WHAT IT MEANS
ACTION
22
23
* See the user instructions (video) at ht-one.nl/support or read the procedure to be followed
E-b
E-U
E-t
E-F
E-0,
E-A,
E-C,
E-E
Appears when the batteries are
almost drained.
Appears when a used test strip
is inserted or a test strip is
incorrectly used *
Appears when the temperature
is higher or lower than the
system's operating range.
Appears when the test strip is
removed during the countdown
or when there is not enough
blood present.
There is a problem with the
meter.
Immediately replace the batteries.
Repeat the test with a new test strip.
The system's operating range is
between 10 °C and 40 °C (50 °F and
104 °F). Repeat the test when the
meter and test strip are within that
range.
Read the instructions and repeat the
test using a new test strip. If the issue
continues, contact your local customer
support for assistance.
Repeat the test using a new test strip.
If the meter continues to malfunction,
contact your local customer support for
assistance.