This test using control solution is not required when first using the  TD-GLUCO 

Bluetooth blood sugar control system.

 

Performing a test with the control solution

When should you perform a test with control solution?

•  When you think that either the meter or the test strips are defective.

•  When your test results do not match the way you are feeling, or when you believe  

 

the results are not accurate.

•  When practicing the test procedure.

•  If you have dropped the meter or have reason to believe it is damaged.

 

Test strips 

(c)

, control solution 

(d)

, lancing device 

(e)

 or sterile lancets 

(f)

 might not be 

included in the kit (please check the list of contents on your product's packaging). These 

articles can be ordered separately.

Performing a test with the control solution.

In order to perform a test with the control solution, you will need 

(b)

(c)

 and 

(d)

.

1.   Insert the test strip into the meter to activate it.

Wait until the meter displays " " and a flashing " ".

2.   Apply the control solution (g)

Shake the bottle of control solution well before use. Squeeze out the first droplet and 

wipe it off. Squeeze another droplet onto the point of the bottle's cap. Hold the meter so 

that the absorbent opening of the test strip comes into contact with the drop and absorbs 

the control solution. Once the test strip's control window contains sufficient control 

solution, the meter starts an automatic countdown.

QUALITY CONTROL TEST

12

13

3.  Reading and comparing the results 

Once the countdown to 

0

 is completed, the result of the test using control solution is shown 

on the display. Compare the result with the range stated on the test strip packaging. 

 The result must be within this range of values. If that is not the case, re-read the instructions 

and repeat the test with control solution.

NOTE:

•  The range for the control solution that is stated on the test strip tube is only valid

  when control solution is used. It is not a recommended range for your blood 

  sugar levels.

•  See the section under 

MAINTENANCE

 for important information concerning 

  the control solution

WARNING:

When you perform a test using control solution mark it as such so that the test result  

is 

NOT

 confused with the blood sugar 

TEST RESULTS

 in the device's memory. If you 

do not do this the blood sugar test results will be mixed up with the test results in 

the device's memory.

Summary of Contents for GLUCO TD-4277B

Page 1: ...BLOOD SUGAR CONTROL SYSTEM User Manual TD 4277B version 2 0 2016 03 HT102015B 01G 311 4277200 002...

Page 2: ...hould you have any further queries after reading this manual please contact your local customer support or your distributor Use This system is intended for exterior use in vitro diagnostic use in a no...

Page 3: ...at could mean incorrect results 9 Do NOT use the meter near sources of strong electromagnetic currents as the accuracy of the measurement could be affected 10 Proper maintenance is essential to the op...

Page 4: ...ue to receive results that are lower or higher than normal then contact your GP and follow their advice Only use fresh whole blood samples for testing your blood sugar levels Using any other substance...

Page 5: ...ld be incorrect if the contact point is not fully inserted into the test slot NOTE The TD GLUCO Bluetooth meter may only be used with TD GLUCO TD 4360 test strips Using other test strips with this met...

Page 6: ...o switch between the General AC PC and QC modes Before using your meter for the first time or whenever you replace the battery check and adjust the following settings Launching the setup mode a Start...

Page 7: ...wipe it off Squeeze another droplet onto the point of the bottle s cap Hold the meter so that the absorbent opening of the test strip comes into contact with the drop and absorbs the control solution...

Page 8: ...full when enough blood is applied Do NOT take your finger away before the device beeps WARNING In order to limit the chance of an infection Never share a lancet or lancing device with others Always us...

Page 9: ...ance with the local regulations 16 17 The meter stores the 1 000 most recent blood sugar test results together with their dates and times In order to access these records start with the device turned...

Page 10: ...unique four digit letter combination of the Bluetooth MAC associated with your meter 4 Add your meter press and press Yes to save the settings 5 Once the devices are linked the data in your meter wil...

Page 11: ...nlight and heat Reseal the tube immediately after removing a test strip Ensure that your hands are clean and dry before touching a test strip Battery Your meter comes with two 1 5 VAAA alkaline batter...

Page 12: ...order to guarantee a pure sample and accurate result Keep the control solution properly sealed at a temperature between 2 C and 30 C 35 6 F and 86 F DO NOT freeze If you perform the recommended actio...

Page 13: ...upside down or not full inserted Meter or test strips are defective Insufficient blood sample Defective test strip Sample has been applied after the device automatically switched off 5 minutes after...

Page 14: ...Temperature warning Operating conditions 10 C to 40 C 50 F to 104 F under 85 relative humidity non condensing Meter storage transport conditions 20 C to 60 C 4 F to 140 F under 95 relative humidity T...

Page 15: ...BorkstraBe 10 48163 M nster Germany Import Europa HTMedical BV Distributor GD Medical Pharma BV Hastelweg 224 5652 CLEindhoven info gdmedicalpharma nl TheBluetooth wordmarkandlogosareregisteredtradema...

Reviews: