![TASLER LTT500 Manual Download Page 35](http://html.mh-extra.com/html/tasler/ltt500/ltt500_manual_821036035.webp)
5. Channel setup
Operation Manual Version 1.6
35
5.2.5. Strain-Gauge Fullbridge
Strain-Gauge Fullbrige is named in this software with DMS Fullbridge (short DMS: DehnungsMessStreifen,
German). It is a fullbride with constant voltage supply. The pin layout is like following:
Connection type
Pin Layout
Description
Strain-Gauge-Fullbridge
Pin 6: Power Out +
Pin 5: Sense +
Pin 3: Input +
Pin 2: Input –
Pin 7: Power Out -
The differential input signal of this channel will be outputted on its output. The amplification or attenuation
depends on the selected input range in the next column. With the range ±5V (1 : 1, or 0dB) the output gives the
input signal 1 : 1 out. Following input ranges can be selected: