15
2.4.2
Operation
Switch on the key-operated mains switch. The following will now occur:
a.
Power ON lamp will be lit
b.
HV OFF lamp will be lit
c.
The variable overload reset lamp will be lit
Press the variable overload reset push button. This will arm the trip circuit.
Select the desired voltage range on the kV meter and set the desired overload trip level.
Before commencing testing ensure that all external interlocks are closed, the emergency stop
button is released and that the regulator is in the fully anti-clockwise position.
Press the HV ON pushbutton. The HV output will now energise, and the red HV on indicator
in the pushbutton will light up. The output voltage can now be increased to the desired level as
indicated on the kV meter.
In the event of a test object failure the HV output will be automatically switched off by the
protection system.
When the test is completed, turn the regulator control knob fully anti- clockwise and switch off
the HV output and then the mains supply.
In case of emergency, the emergency stop push button is provided, switching off the HV
output when engaged.