44
[Send Settings]
button
[Initialize] button
Enter Values
Check one
of these
[Close] button
[Adjustment] Dialog Box
4.
Select either [1-point Adjustment] or [2 Point Adjustment].
5.
Please enter the actual pre-adjusted value(s) in the [Current
Reading] column and the post-adjusted reference value(s) in
the [Post-adjusted Reading] column.
6.
Click on the [Send Settings] button to send the adjustment
equation to the selected Data Logger.
ɜ
[Initialize] button:
Clears any previous adjustment equation(s) from the Data Logger.
ɜ
[Send Settings] button:
Transmits the adjustment equation to the Data Logger.
ɜ
[Close] button:
Quits and closes the window.
Ȇ
We cannot guarantee that after carrying out adjustment the measuring accuracy
will improve for all measuring ranges.
Ȇ
By carrying out adjustment on an already adjusted device without fi rst clicking the
Initialize button, the new adjustment equation will be be used to further adjust the
previously adjusted result. It is important to remember that if adjustment takes
place without initializing the device, the adjustment will include the fi rst equation
as well as the second and an over adjustment will occur. This over adjustment will
be especially large when making a 2-point adjustment and the measurement falls
outside of the range of the two points entered.
When you are carrying out adjustments and formulating equations please take
care with the settings and take personal responsibility for the process.