42
RTR-51A/52A: Value Adjustment Function
Currently the adjustment functions can be used only with the
RTR-51A, 52A. This function enables the user to make adjustments
to measured values to correct inaccuracies that may occur when
compared to a standard or reference temperature measurement. By
setting an adjustment equation beforehand, the adjustment function
will automatically calculate to record and display the post-adjusted
measurement. It is possible to select from two methods of
adjustment: 1-point adjustment and 2-point adjustment.
¦
When you are carrying out adjustments and formulating equations please take
care with the settings and take personal responsibility for the process.
Ⱥ
Temperature: 1-point Adjustment
Ȼ
Adjustment is carried out by
entering the value of the difference
between the displayed temperature
measurement reading and the
reference one to which it is being
compared.
Please set so that the possible
adjustment range is within
Ⱦ
30
Ɏ
.
Ⱥ
Temperature: 2-point Adjustment
Ȼ
Adjustment is carried out by
entering values for the differences
between two displayed temperature
measurement readings and the
reference ones to which they are
being compared.
Please set so that the possible
adjustment range is within
Ⱦ
3