Stir:
Frequency of rotation change.
Input value from 0-990.
Number “20” means the spin direction changes every 20 seconds.
Procedure:
1.
Press enter/mixing trigger to select the target program
2.
In Cap. & Temp. Edit page, user can change the volume of each well and set a
temperature for whole program by entering a value and ticking in enable box.
3.
Press up key to start editing the details of each step.
4.
Press enter/mixing trigger at name field, you can rename the program.
5.
Press rod button to switch input keys, and press enter/mixing trigger to enter.
6.
Use left and right keys of click wheel to move the cursor.
7.
Use up and down keys of click wheel to adjust the factor.
8.
Set the step at Step field.
9.
Set the action at Mode field.
10.
Set the target well at Well field.
11.
Set the period of the step at Time field.
12.
Set the temperature value at Temperature field, please tick in the enable box
for temperature control.
13.
Stir field indicates the frequency of rotation change. Number “20” means the
spin direction will change every 20 seconds.
14.
Adjust the mixing speed at RPM field.
15.
Move the cursor to save button and press enter/mixing trigger to save the
change.
NOTE: The temperature control only functions when the M8-H connects to the M8-H
Autostage