15
11
Manual Mode
User can perform each operation separately in manual mode.
Mixing
1.
Mount the spin tips to the M8-H.
2.
Insert the M8-H to the bottom of wells and press enter/mixing
trigger to start mixing.
3.
Press enter/mixing trigger to stop mixing.
4.
Use up or down keys to accelerate or decelerate of spin speed.
5.
When the spinning is stopped, use left and right keys to change
the spin direction.
Collection
1.
When mixing stopped, raise the M8-H off the wells and press
rod button to extend magnetic rods.
2.
After the rods are completely extended, slowly insert the M8-
H into the wells. The beads will be collected on the spin tips.
3.
Raise the M8-H off the wells, slowly place the M8-H with beads
collected on the spin tips into the target wells.
4.
Submerge the spin tips with collected beads in solution and
press rod button again to withdraw the magnetic rod. The
beads will be released in the target wells.
Summary of Contents for Maelstrom 8 Autostage
Page 1: ...Magnetic Rotary Mixer Maelstrom 8 channel Handler...
Page 6: ...4 2 Equipment List...
Page 7: ...5 3 Overview...
Page 8: ...6...
Page 32: ...30 19 3 Overview Main operation interface Power input...