2. Installation
2.1 Installation Site
Install the Tester on a draft-free, level and stable table. The Tester needs to be placed at least 20cm
away from other objects.
The following locations should be avoided as the installation site:
- An area with high temperature and/or high humidity
- An area exposed to direct the sun
- An area with poor ventilation and/or dusty air
- An area subject to rapid changes in ambient temperature
2.2 Electric Connections (Fig. 3)
WARNING To prevent electric shock hazard, be sure the grounding wire is
connected.
Do not ground to gas pipe; fire or explosion may result.
The grounding wire may be connected to one of the following:
- The ground terminal of the power outlet
- Copper wire buried at least 65cm deep in the earth
- Metallic water pipe line appropriate for use as electric
grounding
(1)
Confirm the voltage of the serial number label located left side of Control Unit
and your electric source.
(2)
Connect the Control Unit and the Test Unit with Signal Connecting Cable.
(3)
Connect the power connecting cable of Control Unit to the Test Unit.
(4)
Plug in the AC power code of the Control Unit to an electric outlet. Be sure to get grounding to
prevent an accident.
2.3 Gas Connection
WARNING The maximum gas pressure should be lower than 9.8 kPa.
Using the rubber hose supplied for gas, connect the gas source (LP gas or Natural gas) and the gas
inlet nipple on the rear panel of the Test Unit. Then fasten both ends with hose bands. See
Fig. 3
.
!
!