18
4. Data Storage Function (use this function with optional Built-in Clock Board)
ACO-7 has a data storage function. Up to 50 successive data can be memorized. (
*4
)
After ACO-7 detects a flash point, the data will be stored automatically.
If Built-in Clock Board is not installed, actual date will not be memorized in
DATE
column.
The recalling procedure is as follows.
Press the
MODE
switch and
RESET
switch, then consecutively release the
RESET
switch and
then the
MODE
switch. The following screen will appear:
Fig. 10 Data Recall Screen
By pressing Up-Down key, the screen will be scrolled.
If needed, press
FUNCTION
key to print out all data. (when optional printer is connected)
*4
: Following data will not be stored.
-
used fire mode, special mode and user custom mode
-
flash is detected at first application
-
flash is detected by manual application
-
(New Data)
(Old Data)
ACO DATA SAMPLE ID FLASH.P
01 03/14 09 LUBE 01 244.0
o
C
02 03/13 18 ST-AS 312.0
o
C
03 03/13 17 BL-AS 338.0
o
C
Month
Day
Hour
Sample ID
Corrected
Flash Point
Number
01: latest
50: oldest