17
3.6 Test by Fire Mode
WARNING Fire Containment Lid and pivot shaft be kept clean to function
properly. Do the damper check before testing by fire mode.
As some fire may not be extinguished completely by the Fire
Containment Lid, the operator may have to extinguish the fire
manually. Prepare fire-resistant gloves and other necessary
objects to protect against burns.
Use knob when resetting the fire containment lid.
The following shows only differences from Normal mode. See Section
3.4.1
to
3.4.6
also.
(1)
When the flash point is detected, the buzzer beeps and the RUN lamp flashes. At this point, test
flame application at the 2
o
C (5
o
F) interval is suspended to prevent accidental fire. Pressing the
START
switch while the
RUN
lamp is flashing will stop the buzzer, and the test is continued
until the fire point detection. If the
START
switch is not pressed within one minute, the test is
terminated and the cool down cycle is initiated.
(2)
Once flash detector detects 5 seconds of continuous burning (or press
START
switch by visual
determination), the fire containment lid activates automatically to help putting out fire. The
flash point and fire point is displayed (flashing alternately every 5 seconds), thus concluding
the test cycle.
(3)
The operator must confirm that the fire containment lid has extinguished burning inside the
cup.
(4)
The operator press
COOL
switch to start the sample cooling process.
(5)
The operator carefully reset the fire containment lid after cooling is finished. Please use caution
resetting the fire containment lid as it possibility may be hot to touch. To reset, push the knob
straight back until latches and turn lid up.
3.7 Test by Skim Mode
The following shows only differences from Normal mode. See Section
3.4.1
to
3.4.6
also.
(1)
To change start temperature and interval of skimmer, press the
FUNCTION
switch to change
the screen, then press the
Right-Left
key to move the cursor, and after that, operate the
Up-Down
keys.
(2)
Press the
RESET
switch to return the main screen after setting of skimmer.
(3)
Attach a paddle to skimmer shaft.
(4)
Automatic surface film skimmer works to when the setting temperature is reached.
Fig. 10 Skimmer Setting Screen
Skimmer can be worked not only automatically but also manually by pressing
MODE
key during a
test.
ACO Baro(hPa) SKIM
1013 Start Interval
300
o
C 10
o
C
Prev.FP 137.0
!