Van ’t Hoffstraat 12
2665 JL Bleiswijk, The Netherlands
T. 31 (0) 10 522 43 73
tv2000-4000mkI-4000DC.docx Rev. 1.06 UK 0718
E-mail: [email protected]
Website: www.tamson.com
Tamson Instruments bv
Page 23/33
ISO 9001 : 2015
NL/PRO 238239125
VAT: NL 80 66 34 984 B01
Bank account no.:
NL28 INGB 0007 350 370
NL95 RABO 0160100046
Chamber of commerce 27 16 95 41
8.3
Stability TV17000
8.4
Influence ambient temperature TV17000
Probe A Probe B
Max
21,699
21,708 °C
Min
21,680
21,688 °C
0,019
0,020
°C
±
0,010
0,010
°C
stDev
0,003848 0,003843
°C
Duration
[hh:mm]
Stability bath Probe A & B @ 150mm
02:05
Δ
Probe A Probe B Ambient
Max
21,700
21,707
27,573 °C
Min
21,618
21,627
23,168 °C
Δ
0,082
0,080
4,405 °C
±
00:59
00:57
2,203 °C
stdev
00:38
00:37
°C
Duration
hh:mm
Stability over 7:42 hrs
07:48
Probe A Probe B Ambient
Max
21,628
21,636
23,344 °C
Min
21,618
21,628
23,168 °C
Δ
0,010
0,008
0,176 °C
±
0,005
0,004
0,088 °C
stdev
0,002156 0,001664
0,0437 °C
Duration
[hh:mm]
Stability when ambient Δ<0,2°C
01:28