Van ’t Hoffstraat 12
2665 JL Bleiswijk, The Netherlands
T. 31 (0) 10 522 43 73
tv2000-4000mkI-4000DC.docx Rev. 1.06 UK 0718
E-mail: [email protected]
Website: www.tamson.com
Tamson Instruments bv
Page 9/33
ISO 9001 : 2015
NL/PRO 238239125
VAT: NL 80 66 34 984 B01
Bank account no.:
NL28 INGB 0007 350 370
NL95 RABO 0160100046
Chamber of commerce 27 16 95 41
4.7
Bath Fluids and use
We recommend the use of the following liquids for the
respective ranges:
Recommended bath fluids
Range
Ordering code
Description
Ambient to
80
C/176
F
N.A.
Clean tap water (preferably de-carbonized)
Do not use DISTILLED. Filtered water by reversed osmoses can
be used without problem.
80..150
C /
176..302
F
00T0220 (20 ltrs)
Tamson mineral oil 150. Transparent, 4 mm
2
/S @ 80°C/176°F; 3
mm
2
/S @ 150°C/302°F; F.P. 200°C/392°F.
20..150
C /
176..302
F
08T0001 (20 ltrs)
Silicon oil 200-10, transparent. 10 mm
2
/S @ 25°C/77°F; 2 mm
2
/S @
150°C/301°F. F.P. 315°C/599°F. When not polluted lifetime is
unrestricted.
150..230
C /
302..446
F
00T0238 (20 ltrs)
Silicon oil 550, transparent. 150 mm
2
/S @ 50°C/122°F; 6 mm
2
/S @
150°C/302°F. F.P. 315°C/599°F. Lifetime at 200°C/392°F is 1year, at
250°C/482°F 1200hrs (when not polluted)
The oil used has a limited lifetime.
The type brand and operating temperature mainly
determines lifetime. Spilling of sample may also reduce
lifetime and in some cases can start chemical reactions.
Silicon oil has the tendency to form gel, for this reason
silicon oil has to be replaced as soon as visible changes
like string forming are noticed. Within a few hours silicon
oil can transform itself into solid gel, which is very difficult
to remove. When not totally cleaned, very small pieces of
left over gel will catalyse new oil to form gel!
Do not use distilled water. This will cause corrosion
and wear of stirrer bearings.
Only use water as a bath fluid below 80°C/175°F.
Working for a longer period with water at
temperatures above 80°C will damage the stirrer
bearings.
When working at temperatures below 10°C / 50°F
ethylene-glycol should be added to the water. For
example, in the volume proportion of 50 / 50%.
CHECK
EXPIRY
DATE
NO
TEMPERATURE
> 80ºC
IF
TEMPERATURE
< 10ºC THAN
ADD ETHYLNE-
GLYCOL
Oil
H
2
O
H
2
O