4.3
Bath Liquid
The bath must be filled with a liquid suitable for the
minimum operating temperature.
It is very important to select a liquid with a viscosity
preferably less (e.g. 10 mm
2
/S) at the operating
temperature and a flash point which is well above the
operating temperature.
The use of other liquids is allowed as long as the viscosity
of the fluid is low enough at the operating temperature.
Following products can not be used as bath fluid:
-
Distilled water,
-
Aggressive fluids when in contact with stainless
steel 304, 316, glass, brass, and silicon sealing.
The fluid flash point must be well above the maximum
operating temperature.
4.4
Drainage of Bath Liquid
The bath can be emptied via the drain tap located at the
backside of the apparatus. For safety reasons the tap can
only be opened by using a screwdriver. The thread inside
the tap is 3/8".
Take necessary precautions against fire hazard when
removing flamable bath fluid.
When removing bath fluid do not inhale toxic vapor.
Always use appropriate ventilation.
Handle old bath fluid as toxic waste.
Cool down bath fluid to ambient before removing.
VISCOSITY
< 10 mm
2
/S
FLASH POINT
> OPERATING
TEMP.
USE
SCREWDRIVER
TO OPEN DRAIN
3/8" BSP inside
thread