Safety Precautions (Must be Obeyed)
WARNING
Insert the chair power plug directly into an outlet by itself (do not share
an outlet with other plugs).
Use an unshared 230-V AC (7 A or more*) outlet with an attached ground terminal.
Using the same outlet together with the power plug of other devices can result in abnormal heat
generation and possible fire.
* 220-V AC (7 A or more), 240-V AC (7 A or more), 120-V AC (12 A or more), or 110-V AC (12
A or more)
Secure insert the power plug into the outlet.
Be sure to fully insert the power plug into the outlet until reaching the base of the plug. Failure
to fully insert the plug or similar fault can result in a short circuit or heat generation that causes a
fire.
Be careful of power cord or
plug breakage.
Do not use any power cord or plug that is
damaged, or that fits loosely when
inserted into an outlet. Doing so can
result in electric shock or fire.
Unplug the power plug when
the chair is not in use.
Turn off the power switch and remove the
power plug from the outlet when the chair
is not in use. Unexpected operation can
result in an accident or fire from a short
circuit due to insulation aging.
Securely connect to ground.
Failure to make a ground connection can
result in electric shock if there is a
malfunction or short circuit. Consult an
electrician for grounding procedures.
Do not wash the chair by
pouring water on it.
Do not wash the chair by pouring water
on it because this chair is an electrical
device. Washing in such a manner can
result in a malfunction or electric shock.
It is absolutely prohibited for
any person other than
qualified technical service
personnel to disassemble or
repair this chair.
You must never attempt to repair or
modify this chair by yourself. Doing so
can result in electric shock during
disassembly or repair, and an improper
repair can cause a fire or abnormal
operation that results in injury.
Be careful to keep hands,
fingers and other body parts
of the customer and user from
touching or getting close to
moving parts.
There is a danger of pinching and other
physical injury.
Hold by the power plug when
unplugging.
Do not pull the cord to remove the power
plug from the socket. Hold by the power
plug at the end of the cord when
removing. Removing the plug by pulling
the power cord can damage the power
cord, resulting in electric shock or fire.
Do not stand on the edge of
the step.
Do not stand on the edge of the step.
Doing so could cause the chair to tip over
or the step could suddenly move,
resulting in injury or damage.
Do not sit on the backrest of
the chair.
Sitting on the backrest of the chair could
cause it to tip over, resulting in injury or
damage.
Do not ground the chair to the
following types of locations.
Water Pipes
Complete grounding is not possible as
hard plastic pipes are widely used.
Gas Pipes
Connecting to gas pipes can result in an
explosion or fire.
Lightning rods and telephone lines
Summary of Contents for LUAR AB-MF