DXF-200 Getting Started Guide
Page 11
INTERNAL INTERLOCK 2
This internal interlock checks to make sure the metal covers
are in place for the high voltage power section of the instru-
ment. When the covers are closed the INTERNAL INTER-
LOCK 2 LED changes from red to green. When all interlocks
are green, the flashlamp is safe to fire.
UNLABELED INTERLOCK 3
This function is not used. This indicator will remain green.
UNLABELED INTERLOCK 4
When the internal interlocks for the optical path are enabled
and the operating software enables it, the UNLABELED
INTERLOCK 2 LED changes from red to green. When all
interlocks are green, the flashlamp is safe to fire.
POWER
When the flashlamp 12 VDC power supply is ON, the green
POWER LED lights up.
INTERLOCK
This function is not used.
CHARGE
The flashlamp controller has 4 control states. The CHARGE
indicator shows when the controller is charging the main
capacitors with the programmed flash voltage.
ARM
The flashlamp controller has 4 control states. The ARM indi-
cator shows when the controller has charged the fire capacitor
and is able to fire the flashlamp.
FIRE
The flashlamp controller has 4 control states. The FIRE indi-
cator shows when the controller has fired the flashlamp.
DUMP
The flashlamp controller has 4 control states. The DUMP
indicator shows when the controller has dumped all the
charge on the main capacitor and disabled the fire circuit.
Table 1:
DXF Front Panel LED Indicators - Flashlamp
Indicator
Function