6.10
"Pressure Loss Measurement" program
Starting the "Pressure Loss Measurement" program
When you start the "Pressure Loss Measurement" program, the
pressure sensor is first zero-balanced; this takes a few seconds.
The colour of the pressure value then changes from grey to black.
The black colour indicates that the device is ready for
measurements. Zero balancing can also be performed manually via
the main menu.
Connecting the system to S4600-ST
®
.
The system to be measured must be connected to S4600-ST
®
. The
pump is then used to apply the test pressure to the system as per
instruction.
"START" starts the pressure loss measurement. A counter in the
main display shows the duration of the running pressure loss
measurement in seconds and minutes.