Surgical Technique USS™ Universal Spine System DePuy Synthes 11
Rod contouring
Use the Rod Template for USS Rods Ø 6.0 mm
(388.870/880) to determine the proper rod contour and
length.
Contour the rod using either the Bending Pliers with
Rolls (1) (388.960) or the USS Bending Irons (2) (388.910
left, and 388.920 right).
Warning:
Do not bend titanium rods backwards and
do not bend rods more than 45°.
If necessary, the construct can be extended by connect-
ing two rods with a parallel or extension connector.
Note:
Hook/screw offset.
Anatomical conditions sometimes result in the implants
not being aligned in a straight line. The screws and
hooks have a 4-mm offset. If the implants are not
aligned during rod insertion it may be necessary to
rotate the screw by 180 degrees or choose a different
hook (i.e. left or right side opening).
1
4 mm
2