Page 6 of 17
Figure 3:
Push button on the receiver
Figure 4:
Push button on the hand control
5.
Start operation: Unlock the hand control, grab the battery pull tab and remove it by pulling in the direction
indicated by the arrow and press button on the hand control to start handpiece. Increase pressure on button to
increase speed of handpiece and release pressure on button to decrease speed of handpiece.
Warning:
Keep hand control safety cover in locked position when not in use.
Figure 5:
Hand control cover shown in unlocked position
6.
Stop Operation: Lock the wireless hand control and unpair the device. To unpair the device simply press receiver
button or unplug receiver from console.
Warning:
Do not re-sterilize or reuse hand control: reuse and/or re-sterilization of hand control could result in
reduced performance, material degradation, and/or patient contamination.