Page 12 of 17
Atmospheric
Pressure
Operating: 0.7
–
1.06bar
Operating: 0.7
–
1.06bar
Transportation and Storage:
Not applicable
Transportation and Storage:
Not applicable
Receiver for Wireless Hand Control, EG1A powered through Console
(SC3001/SC3002) Specifications
Primary: 100-240VAC, 50/60 Hz, 250 VA
Class I: Protective Earth
Wireless Hand Control, EG1A Specifications
Powered through: CR1620 lithium coin battery, nonreusable
–
disposable
Fluid Ingress Protection: IPX2
Guidance and manufacturer’s declaration –
electromagnetic emissions
The
eG1 System
is intended for use in the electromagnetic environment specified below. The customer or the user of the
eG1
System
should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment
–
guidance
RF emissions
CISPR 11
Group 1
The eG1 System uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The eG1 System is suitable for use in all establishments other
than domestic establishments and those directly connected to
the public low-voltage power supply network which supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacturer’s declaration –
electromagnetic immunity
The
eG1 System
is intended for use in the electromagnetic environment specified below. The customer or the user of the
eG1
System
should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment
–
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
8 kV contact
15 kV air
8 kV contact
15 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for AC lines
1 kV for I/O lines
100kHz Pulse Repetition
Frequency
2 kV for AC lines
1 kV for I/O lines
100kHz Pulse Repetition
Frequency
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
0.5, 1kV line to line
0.5, 1, 2 kV line to earth
0.5, 1kV line to line
0.5, 1, 2 kV line to earth
Mains power quality should be
that of a typical commercial or
hospital environment.