Note: DePuy Synthes has not tested compatibility with devices provided by other manufacturers and assumes no
liability in such instances.
Cleaning and Sterilization
Receiver for Wireless Hand Control, EG1A (RECEIVER-HC)
Receivers are non-sterile capital equipment and shall always be maintained outside the Operating Room Sterile
Field.
Warning:
Do not immerse or sterilize the receiver.
1. Disconnect receiver from console.
2. Clean receiver by wiping with non-abrasive cloth and neutral pH detergent and deionized, distilled, or purified
water after each case.
3. Dry thoroughly with non-abrasive cloth.
Wireless Hand Control, EG1A (WIRELESS-HC)
Warning:
Do not reprocess the Hand Control. Hand Controls are disposable and SINGLE USE ONLY.
Reprocessing hand controls may result in loss of function causing delay in surgery.
The Hand Control is provided sterile. Remove product from the package using established operating room
procedures for sterile transfer.
Store sterile devices in their original protective packaging, and do not remove them from the packaging until
immediately before use.
Prior to use, check the product expiration date and verify the integrity of the sterile packaging. Do not use if the
package is damaged.
Explanation of Symbols Used
General Symbols
Direction of motion
Reference or Catalogue Number
(Ref. 5.1.6 ISO 15223-1)
Consult Instructions for Use
(Ref. 5.4.3 ISO 15223-1)
Materials
Caution: United States Federal law
restricts this device to sale by or on
the order of a physician or other
licensed health-care provider
(21 CFR 801.109)
Date of Manufacture
Manufacturer (Ref. 5.1.1 and 5.1.3
ISO15223-1)
Product is ETL listed to the
requirements of US and Canada
Keep dry (Ref. 5.3.4 ISO 15223-1)
Locked
Do not use if package is damaged.
(Ref. 5.2.8 ISO 15223-1)
Unlocked