Page 1: ...Instructions for Use Universal Battery Charger...

Page 2: ......

Page 3: ...of Product Operating Instructions Troubleshooting 2 4 5 Operating the Universal Battery Charger 6 Cleaning 6 Charging the battery 6 Charging a new battery or one which has not been used for a prolong...

Page 4: ...Atmospheric 500 1060 hPA 500 1060 hPA pressure 0 5 1 06 bar 0 5 1 06 bar Technical data Operation voltage 100 VAC 240 VAC 50 60 Hz Max current consumption 1 2 A 530 600 2 4 A 530 601 Mains input powe...

Page 5: ...it switches the drive off automatically when the charge has dropped to the defined residual capacity Only original Synthes products are used these are used properly and the guarantee period has not ex...

Page 6: ...the battery may not be disposed in the normal waste as it contains Metal Hydride Dispose of or recycle in accordance with lo cal state and federal regulations According to WEEE 2002 96 EC the Universa...

Page 7: ...bays for Synthes batteries Charge control indicators red yellow green Symbols to assist operation Main switch Fuses 2 5 AT 250 V SPT 5 20 mm Mains cable connector Ventilation openings Description of...

Page 8: ...mpletely charged the charger switches over to maintenance charging green indi cator lights up The battery can be removed from the charg ing bay at any time but it is only fully charged when the green...

Page 9: ...down and is automatically resumed after a maximum of 25 minutes Red indicator lights up The battery is defective and must be replaced No indicator lights up on charger The charger has no mains power o...

Page 10: ...charger on again 4 Have charger checked and if necessary repaired by the official Synthes manufacturer or an authorised sales outlet Have charger checked and if necessary repaired by the official Synt...

Page 11: ...l down in the charging bay After a maximum of 45 minutes the yellow indicator also lights up After a further 25 minutes the charging process is automatically resumed If the yellow indicator does not l...

Page 12: ...in charger or battery Replace battery Replace battery Replace battery Replace battery Have charger and battery checked and if necessary repaired by the official Synthes manufacturer or an authorised...