BACKUP INDIFFERENT ELECTRODE
INTERCONNECT
The troubleshooting section of this manual may tell you to use the
backup indifferent
electrode interconnect.
This is called the
backup connector
for short. It is part
number 22-0026 of the patient kit. The backup connector may be useful when the
indifferent electrode under the skin has failed. This is when the NeuRx
®
EPG displays
“XXXX” (see Figure 25 below) but the patient cable and EPG still work correctly.
Figure 25
. EPG displaying “XXXX”
.
Follow these steps to use the backup connector:
1. Disconnect the patient cable from the electrode connector (Figure 26).
Figure 26