35
3. When you are done bathing, carefully uncover the wires and the connector
holder.
4. If you notice moisture on the electrode connector, wipe it with an alcohol wipe.
Use a wipe with 70% isopropyl alcohol. Allow the alcohol to air dry before use.
CLEANING OF COMPONENTS
CAUTION: Do NOT soak the NeuRx
®
EPG or the patient cable in liquid
.
Clean the NeuRx
®
EPG and the patient cable as needed. Wipe the outside of the
NeuRx
®
EPG and the patient cable using a clean, damp cloth with an anti-bacterial
cleaning solution. Be sure to wipe away any leftover cleaner.
TROUBLESHOOTING – NeuRx
®
EPG
If you think that the NeuRx
®
EPG is not working properly or multiple “X”s appear on
the screen, please follow the steps below:
1. Check the connection between the patient cable and the NeuRx
®
EPG. Check
the connection between the patient cable and the electrode connector. Make
sure these connections are secure.
2. Replace the patient cable with the spare cable in the patient kit.
3. Disconnect the patient cable from the NeuRx
®
EPG.
4. Insert the test plug.
a. When the test plug is inserted into the NeuRx
®
EPG, the screen should
show “AAAA” when activated.
b. If “AAAA” does not appear on the screen or if there is still a problem, call
Synapse Biomedical.
For help with your NeuRx
®
DPS device, first call your healthcare provider who is
helping you with your NeuRx DPS
®
. As a back-up you may also call:
Synapse Biomedical Customer Service
U.S. Toll Free: 1-888-767-3770
From outside the U.S.: 001-440-774-2488