Chapter 2 Performance Tests
58502A Operational Verification
2-10
User’s Guide
4
Set the RS-232 port of your PC to match the following default values:
If needed, see “Connecting a Terminal or Computer to the 58502A”
starting on page 1-5 in Chapter 1 for information on setting above
parameters.
a.
Press Return (or Enter) key on the PC.
The
scpi>
prompt should be displayed.
b.
At the prompt, type
*IDN?
c.
Press Return.
This query returns the 58502A identification similar to the
following:
HEWLETT-PACKARD, 58502A,3426A–00123,3422 – A
5
Mark Pass or Fail in Line 3 on the Operational Verification portion of
the 58502A Performance Test Record, located on page 2-14.
Pace:NONE
Baud Rate: 9600
Parity: NONE
Data Bits: 8