Chapter 2 Performance Tests
58502A Calibration (Option 010 Only)
2-18
User’s Guide
Figure 2-7. 58502A Calibration Setup
4
Observe Counter reading.
If difference from 10 MHz is
<
5
×
10
−
8
(0.5 Hz), then perform the
“Software Oscillator Adjustment” procedure below.
Software Oscillator Adjustment
1
Use the terminal and serial port to send the following SCPI command
to ascertain the EFC value currently applied:
DIAG:CAL:ROSC:EFC:ABS?
HP 53131A
Universal
Counter
or equivalent
BNC Cable
BNC Cable
Outputs
1 - 12
HP 24542U
Interface
cable
or equivalent
RS232C
Port
To any
available
Output
A
B
1
2
3
4
5
6
7
8
9
10
11
12
5 VRMS MAX
100-240 VAC 50/60GHz
40 VA Max
!
!
!
58502A
Distribution
Amplifier
Terminal or PC
Ref. In
50
Ω
Feedthrough
(HP 10100C)
5071A
PRIMARY FREQUENCY STANDARD
Continuous
Operation
HP 5071A
Primary Frequency
Standard or equivalent
Channel 1
Port1
10 MHz External
Reference Out
Summary of Contents for 58502A
Page 10: ...In This Guide x User s Guide ...
Page 11: ...1 Getting Started ...
Page 27: ...2 Performance Tests Verifying Specifications ...
Page 29: ...Chapter 2 Performance Tests Introduction User s Guide 2 3 ...
Page 47: ...3 Features and Functions ...
Page 59: ...4 Command Reference ...
Page 73: ...5 Functional Description ...
Page 78: ...Chapter 5 Functional Description Functional Block Description of the 58502A 5 6 User s Guide ...
Page 79: ...6 Service ...
Page 81: ...7 Specifications ...
Page 84: ...Chapter 7 Specifications Introduction 7 4 User s Guide ...