Chapter 2 Performance Tests
Test Equipment Required
2-4
User’s Guide
Test Equipment Required
Table 2-1. Recommended Test Equipment
Instrument
Required Characteristics
Recommended Model
Use
*
* C= Calibration (Option 010 only)
OV = Operational Verification
P = Performance Tests
Signal Generator
Frequency Range: 100 kHz to 10 MHz
Output Level: +6 to +16 dBm
Harmonic Distortion:<-30 dBc
Spurious Distortion:<-80 dBc
HP 8663A (or equivalent)
OV, P
Digitizing
Oscilloscope
Frequency range: 100 kHz to 100 MHz HP 54600B (or equivalent) OV
Universal Counter
Frequency Range: 100 kHz to 100
MHz: >10 digits resolution @ 10 MHz
HP 53131A (or equivalent) C (for
Option 010
only)
10 MHz Reference
Source
10 MHz Reference Signal
Accuracy:
±
2 X 10
-12
HP 5071A (or equivalent)
C
Terminal or computer
Communication software, RS-232
connection
Any Model
OV, P, C
RMS Voltmeter
Accuracy:
±
2% at 3V Full Scale,
1 MHz
HP 3400B (or equivalent)
P
50
Ω
BNC
Feedthrough
_____________________________
HP 10100C (or equivalent) OV, P, C
RS-232 Interface
Cable
DTE-to-DTE 9-pin (f) to 9-pin (f)
HP 24542U (or equivalent) C
Type N-to-BNC
Connector
_____________________________
H P 1250-0780 (or
equivalent)
OV, P
50
Ω
BNC Coaxial
Cable (2)
RG-58, 12 meter, male-to-male
HP 10503A (or equivalent) OV, P, C
Summary of Contents for 58502A
Page 10: ...In This Guide x User s Guide ...
Page 11: ...1 Getting Started ...
Page 27: ...2 Performance Tests Verifying Specifications ...
Page 29: ...Chapter 2 Performance Tests Introduction User s Guide 2 3 ...
Page 47: ...3 Features and Functions ...
Page 59: ...4 Command Reference ...
Page 73: ...5 Functional Description ...
Page 78: ...Chapter 5 Functional Description Functional Block Description of the 58502A 5 6 User s Guide ...
Page 79: ...6 Service ...
Page 81: ...7 Specifications ...
Page 84: ...Chapter 7 Specifications Introduction 7 4 User s Guide ...