Configuration Menu
Page 23/28
Programming instructions of the
SYCLOPE INDIG’O
®
Polarization : define the duration before launching the control
Name
Signification
Setting
range
Default
value
Action
Polarization Delay before launching the control 0……15 min
2…min
Stop control
Display picto
6)
Output testing : Page “Test”
This page allow you to check manually the working of the output :
R1 relay : Activation of the powered relay
P
1Dosage
R2 relay : Activation of the dry contact relay
R
1Alarme
Output I : Sending a defined current on the analog output
I
OUT1
Name
Signification
Setting
value
Default
value
R1 relay
Working test of
P
1Dosage
Open or Close
Open
R2 relay
Working test of
R
1Alarme
Off or On
Off
Output I
Working test of
I
OUT1
0…20 mA
0,00
Sensor
Flow switch
2
Polarization :
min
Free Cl
mg/l
1,09
Setpoint : 2,00 mg/l
REG
ALR
Dosage : 0,00%
Output : 5,750 mA
switch
Sensor
TEST
Open
R1 relay :
Off
R2 relay :
0,00
Output I :
mA