Program List and Explanations
5.3.2.6
Signal Outputs
: Select operating mode of the signal output:
5.3.2.7
Output/Control
: Select operating mode of the controller output:
The relay contact can be activated daily, at any time of a day.
5.3.2.341
Start time
: to set the start time proceed as follows:
1
Press [Enter], to set the hours.
2
Set the hour with the [
] or [
] keys.
3
Press [Enter], to set the minutes.
4
Set the minutes with the [
] or [
] keys.
5
Press [Enter], to set the seconds.
6
Set the seconds with the [
] or [
] keys.
Range: 00:00:00–23:59:59
5.3.2.44
Run Time
: see Interval
5.3.2.54
Delay
: see Interval
5.3.2.6
Signal Outputs
: see Interval
5.3.2.7
Output/Control
: see Interval
The relay contact can be activated at one or several days, of a
week. The daily starting time is valid for all days.
Cont.:
Signal outputs continue to issue the measured value.
Hold:
Signal outputs hold the last valid measured value.
Measurement is interrupted. Errors, except fatal errors,
are not issued.
Off:
Signal outputs are switched off (set to 0 or 4 mA).
Errors, except fatal errors, are not issued.
Cont.:
Controller continues normally.
Hold:
Controller continues based on the last valid value.
Off:
Controller is switched off.
5.3.2.24
daily
5.3.2.24
weekly
Summary of Contents for AMI Oxytrace
Page 1: ...AMI Oxytrace AMI Oxytrace QED Version 6 20 and higher A 96 250 531 300617 Operator s Manual...
Page 5: ...AMI Oxytrace QED A 96 250 531 300617 3 11 Default Values 90 12 Index 93 13 Notes 94...
Page 96: ...94 A 96 250 531 300617 AMI Oxytrace QED Notes 13 Notes...
Page 97: ...AMI Oxytrace QED Notes A 96 250 531 300617 95...