54
A-96.250.531 / 300617
AMI Oxytrace QED
Maintenance
1
Stop sample flow to the monitor AMI Oxytrace by closing the
appropriate valve, e.g. back pressure regulator, sample prepa-
ration or flow regulating valve at flow cell.
2
Connect sample line of the monitor AMI Oxytrace [A] with the
sample inlet of the reference instrument AMI INSPECTOR Oxy-
gen [B]. Use the supplied tube, made of PA.
3
Connect sample outlet of the reference instrument AMI IN-
SPECTOR Oxygen to the sample outlet funnel of the monitor.
4
Switch on the AMI INSPECTOR Oxygen. Open the flow regulat-
ing valve and regulate the sample flow to 10 l/h. The actual flow
is shown on the transmitter.
A
B
C
D
Monitor AMI Oxytrace
AMI INSPECTOR Oxygen
Reference flow cell
On-line flow cell
E
F
G
Sample outlet
Sample inlet
T-fitting
A
B
C
D
E
F
G
Summary of Contents for AMI Oxytrace
Page 1: ...AMI Oxytrace AMI Oxytrace QED Version 6 20 and higher A 96 250 531 300617 Operator s Manual...
Page 5: ...AMI Oxytrace QED A 96 250 531 300617 3 11 Default Values 90 12 Index 93 13 Notes 94...
Page 96: ...94 A 96 250 531 300617 AMI Oxytrace QED Notes 13 Notes...
Page 97: ...AMI Oxytrace QED Notes A 96 250 531 300617 95...