AMI Oxytrace QED
Program List and Explanations
A-96.250.531 / 300617
85
5.3.2.342
Calendar
:
5.3.2.342.1
Start time
: The programmed start time is valid for each of the pro-
grammed days. To set the start time see
.
Range: 00:00:00–23:59:59
5.3.2.342.2
Monday
: Possible settings, on or off
to
5.3.2.342.8
Sunday
: Possible settings, on or off
5.3.2.44
Run Time
: see Interval
5.3.2.54
Delay
: see Interval
5.3.2.6
Signal Outputs
: see Interval
5.3.2.7
Output/Control
: see Interval
The relay will be switched via the Profibus input. No further param-
eters are needed.
5.3.4
Input:
The functions of the relays and signal outputs can be de-
fined depending on the position of the input contact, i.e. no function,
closed or open.
5.3.4.1
Active:
Define when the input should be active:
5.3.4.2
Signal Outputs:
Select the operation mode of the signal outputs
when the relay is active:
5.3.2.1
Function = Fieldbus
No:
Input is never active.
When closed
Input is active if the input relay is closed
When open:
Input is active if the input relay is open
Cont.:
Signal outputs continue to issue the measured
value.
Hold:
Signal outputs issue the last valid measured
value.
Measurement is interrupted. Errors, except fatal
errors, are not issued.
Off:
Set to 0 or 4 mA respectively. Errors, except fatal
errors, are not issued.
Summary of Contents for AMI Oxytrace
Page 1: ...AMI Oxytrace AMI Oxytrace QED Version 6 20 and higher A 96 250 531 300617 Operator s Manual...
Page 5: ...AMI Oxytrace QED A 96 250 531 300617 3 11 Default Values 90 12 Index 93 13 Notes 94...
Page 96: ...94 A 96 250 531 300617 AMI Oxytrace QED Notes 13 Notes...
Page 97: ...AMI Oxytrace QED Notes A 96 250 531 300617 95...