AMI Oxytrace QED
Maintenance
A-96.250.531 / 300617
43
3
Unscrew and remove the sensor cap [H] from the Swansensor
Oxytrace G [A].
4
Empty the remaining electrolyte.
5
Refill the sensor cap with fresh electrolyte.
NOTICE:
There is a groove [C] in the thread [B] of the sensor,
where excessive air and electrolyte can escape while screwing
the sensor cap onto the sensor. Hold the sensor in vertical
position, measuring head pointing downwards.
6
Slowly screw the sensor cap onto the sensor to allow the ex-
cess electrolyte to escape without building up too much pres-
sure inside the electrode. Fasten sensor cap tightly.
7
Clean the sensor thoroughly and dry the sensing membrane
with a soft tissue.
8
Switch on power.
9
Let the sensor run-in at air for at least 30 min, better 1 h.
10
Afterwards perform an air calibration.
11
Install the sensor into the flow cell.
12
Select “New Filling” to reset the counter for remaining electro-
lyte, see <Maintenance>/<Service>
A
B
C
D
E
F
G
H
Swansensor Oxytrace G
Thread
Groove
Guard electrode
Anode
Measuring head
Cathode
Sensor cap with membrane
A
B
C
D
E
F
G
H